Paperless Recorders

Monarch DataChart 1250
DataChart 6000

Tired of messing with paper? A new line of paperless recorders is now available for installation on our test chambers.

A maximum sample storage rate of 100 samples per second can be set for both channels allowing for capture of short duration process signals anomalies. CompactFlash cards up to 2 Gigabyte size can be used allowing many data points to be stored over long periods of time.

With intuitive operation, flexible input configuration and large memory capability, the DataChart 1250 is a powerful, compact digital recorder.

Monarch DataChart 1250 Paperless Recorder

Monarch DataChart 1250 Paperless Recorder

Monarch DataChart 6000 Paperless Recorder

Monarch DataChart 6000 Paperless Recorder



DC1250 Specifications

Input Power: Standard

 

9 Vdc +/- 0.5Vdc @ 5VA (depends on external loads) provided by external AC wall transformer (provided)., non-isolated, 100-240Vac 50/60Hz Input impedance: >100k Ohms

Option:

Isolated 12-24 Vdc input power available (not compatible with internal bater pack option below). Measurement Rate:  Up to 100 samples/sec per channel.

Option:

Internal battery pack provides uninterrupted operation and controlled shutdown during blackout. 6Vdc, 2400mAH NiMH - Backup Time: 8 hours typical (depends on external load). Math Function:  Y=mx + b; average, hi peak, low peak.
Output: 2 outputs 5Vdc @ 50mA to power external sensors. Media:  Compact Flash to 2GB
Number of Channels: 2 universal, user selectable. Display:  LCD Graphics, 160 x 80 pixels, black FSTN with white LED backlight. User controlled backlight level and contrast adjust (electronic).
Isolation: 300V AC/DC channel input to chassis ground Display Modes:  Trending (horizontal), Large dual digital readout, mixed mode.
DC Input: Voltage 
Ranges:
0-250mV; 0-1.25V; 0-2.5V, 0-5V; 0-12.5V; 0-25V
User Interface:  5 button keypad (dual function buttons).
  Accuracy: 0.1% of reading
Resolution: 0.025% of full scale
Clock:  Auto leapyear and daylight savings adjustment. Internal battery backup.
  Current 
Ranges:
0-20mA; 4-20mA; 0-50mA; 10-50mA
Relay Output:  Two alarm outputs: 30V 0.5A Form A relays.
  Accuracy: 0.1% of reading excluding 250 ohm external shunt (required) Opto-isolated Input:  One input, 5 to 12Vc activation @ 10mA typical.
  Resolution: 0-25% of full scale. Audible:  Internal beeper (multiple tones).
Thermocouple: Accuracy: 0.3% of full scale (typical). Ambient Temperature Sensor Accuracy: +/-1.5°C (see datasheet for full specs.) Dimensions:  Front panel: 96mm (1/4 DIN) x 152mm (3.78 x 3.78 x 6 inches).
 RTD: Accuracy: 0.3% of full scale (typical)
Resolution: 0.1°C
Internal Current source: 1mA (see datasheet for full specs)
Environmental:  Indoor Use Only
Installation Category II
per IEC 664
Pollution Degree Level II per IEC61010-1
Temperature: -10°C to 50°C operating per IEC 61010-1
Humidity: Maximum relative humidity 80% for temperature up to 31°C, decreasing linearly to 50% relative to humidity at 40°C.
Frequency Input (Hz) Range: 0-10,000 Hz/ 0-600,000 RPM Optional:  USB: Front panel USB 2.0 slave port for data and remote real-time display when using Navigator software.
Speed Input (RPM: Accuracy: Freq: +/-1 Hx; RPM +/1 RPM below 9,999 RPM
Input: Low <1.0Vdc; High >3.0 <12.0Vdc
Pulse Width: 10 microsecond minimum.
   Ethernet: Real time data display, historic data transfer, remote control and recorder configuration when using Navigator Software. Built in web server displays current measured values.


DC6000 Specifications

General
Input Resolution: 0.0015% of full scale, 16 bit unless otherwise stated
Input Impedance: >1 Mohm
Input Channels: 6 or 12 direct input plus 6 additional calculated channels
Maximum Input: 50Vdc
Isolation: Channel to Channel: 350Vdc or RMS AC
                Channel to Chassis: 2000 Vdc or RMS AC
                Isolation category II: Pollution Degree 2
Measurement Rate: 10 times per second on all direct input channels
Common Mode Noise Rejection: >100dB, 50/60 Hz, filter enabled
Normal Mode Noise Rejection: >50dB at 50/60 Hz, filter enabled
Math Functions: Fully programmable +, -, x, /, square root, sine, cosine, tangent, log, 
totalization, powers, averages, conditional logic; AND, NOT, OR, +, >, <, gated timers. 
Can use live channels in calculation. Can define 6 constants and 6 functions per channel.
 
Analog Inputs
DC Voltage: +/-125mV, +/-250mV, +/-500mV, +/-1.00V, +/-3.0V, +/-6.0V, +/-12.0V, +/-24.0V
Accuracy: Ranges to 1V +/-0.06%, Ranges > 1V +/-0.1%
 
DC Current:  4-20mA, 0-20mA, 10-50mA
Accuracy: +/-0.15% using external 50 ohm 0.1% 1/4 watt shunt
 
Thermocouple (Per ITS90)
Resolution: 0.1C, thermocouple burnout detection: Automatic
Reference junction compensation accuracy: +/-2.5 °C (0 °C to 50 °C)
 
Type       Range (°C) Accuracy (°C)    Range (°F) Accuracy (°F)
J              -210 to 0°C +/-1.6 °C              -346 to 32°F +/-3.0 °F
                -0 to 1200 °C +/- 0.7 °C          32 to 2192 °F +/- 1.2 °F
K             -270 to -162 °C +/- 5.0 °C       -454 to -260 °F +/- 9.0 °F
                -162 to -70 °C +/- 2.2 °C         -260 to -94 °F +/- 4.0 °F
                -70 to 1372 °C +/- 0.7 °C         -94 to 2500 °F +/- 1.2 °F
T              -270 to -73 °C +/- 2.5 °C        -454 to -100 °F +/- 4.5 °F
                -73 to 200 °C +/- 1.0 °C          -100 to 400 °F +/- 1.8 °F
E              -270 to -70 °C +/- 2.5 °C        -454 to 94 °F +/- 4.5 °F
                -70 to 1000 °C +/- 0.7 °C        -94 to 1832 °F +/- 1.3 °F
N             -270 to -100 °C +/- 3.0 °C       -454 to 212 °F +/- 5.4 °F
                100 to 1300 °C +/- 0.9 °C        212 to 2372 °F +/- 1.6 °F
S              -50 to 0 °C +/- 2.8 °C              -58 to 32 °F +/- 5.0 °F
                 0 to 1000 °C +/-1.4 °C            32 to 1832 °F  +/- 2.5 °F
                 1000 to 1710 °C +/- 1.1 °C     1832 to 3110 °F +/- 2.0 °F
                 1710 to 1768 °C +/- 1.4 °C     3110 to 3214 °F +/- 2.5 °F
B              0 to 535 °C +/- 4.0 °C             32 to 1000 °F +/- 7.2 °F
                 535 to 1800 °C +/- 2.1 °C       1000 to 3300 °F +/- 3.8 °F
U              0 to 100 °C +/- 0.9 °C              32 to 212 °F +/- 1.6 °F
R              0 to 260 °C +/- 3.0 °C              32 to 500 °F +/- 5.4 °F    
                 260 to 650 °C +/- 1.4 °C          500 to 1200 °F +/- 2.5 °F
                 650 to 1760 °C +/- 1.0 °C        1200 to 3200 °F +/- 1.9 °F
C              0 to 180 °C +/- 2.5 °C               32 to 360 °F +/- 2.5 °F
                 180 to 1200 °C +/- 0.8 °C         360 to 2220 °F +/- 1.5 °F
                 1200 to 2315 °C +/- 3.6 °C       2220 to 4199 °F +/- 6.5 °F
L               -200 to 0 °C +/- 1.1 °C              -328 to 32 °F +/- 2.0 °F
                 0 to 890 °C +/- 0.7 °C                32 to 1650 °F +/- 1.2 °F  

Note: Field calibration improves accuracy to beter than +/- 2 °F and up to 
+/- 0.5°F and ensures compliance with AMS2750D measurement accuracy
requirements for Nadcap accredited suppliers.                   
   
 
RTD
Base Accuracy: 0.2% or 0.5 °C (1 °F). Resolution: 0.1 °C
2 or 3 wire connection. Cable compensation to +50 ohm. Open and short circuit detection.
 
Type                               Range °C                        Range °F
100 ohm Plt. 385            -220 to 850 °C                 -364 to 1560 °F
100 ohm Plt. 392            -180 to 820 °C                -292 to 1500 °F
200 ohm Plt. 385            -220 to 400 °C                -364 to 750 °F
200 ohm Plt. 392            -180 to 400 °C                -292 to 750 °F
100 ohm Ni.                    -70 to 300 °C                  -94 to 570 °F
120 ohm Ni.                    -70 to 300 °C                  -94 to 570 °F
1000 ohm Ni.                  -60 to 209 °C                  -76 to 408 °F
10 ohm Cu.                     -70 to 170 °C                  -94 to 338 °F*
*0.5% +/- 0.5 °C
 
Frequency inputs (2 or 4 channels)
Range: 0 to 5,000Hz all channels, 0 to 10,000Hz 1 channel
Accuracy: 0.005% +/- 1 digit
 
Recording
Recording Rates: User programmable from 10 samples per second to 1 sample every 24 hours
Data Format: Proprietary encrypted format, User file naming
Data Storage Capacity: Data stored in non-volatile RAM and recorded automatically to:
                Removable media types:
                        CompactFlash™ or USB drive to 4 GB
                Internal media type:
                        SD card (secure digital) to 4 GB
File Types:Data files, alarm and event files, configuration files, language files, multiple files of different names on a single disk.
Display  
Type: Color LED backlit Active Matrix TFT Liquid Crystal Display
Size: 5.6 inch diagonal, Resolution: 320 (W) x 240 (H) pixels
Interface: Resistive analog touch screen control
Display Builder: Allows user to create custom displays
Display Modes: Graphic trending (vertical or horizontal), Bar Graphs (vertical or horizontal), Digital Meter (large or small), Alphanumeric Alarm and Event log
Virtual Chart Speed: Programmable from 0.5 inch/hour to 600 inches/hour (10 mm/hr to 10,000 mm/hr)
Display Windows: Time/Date, Graphics (bars, large digital, trends) Disk Status, System Status, Menu Button Bar, Unit Indentification, Alarms/Events
 
 
Communications
Network: 10/100 Base T Ethernet per 802.3, RJ45 connection standard
Servers: Webserver supports http and ftp protocols, OPC server, Modbus over Ethernet server
Serial: Isolated RS485/RS232 Modbus Interface (option)
 
Power
Requirements: 100 to 240 Vac, 50/60Hz. 35 VA max. Optional 24 Vdc +/-15%
Power Fail Protection: Programmed parameters stored in non-volatile memory. Clock battery backed. Internal battery backup provides orderly shutdown and the ability to survive brownouts and short blackouts (<20 seconds)
Power Output: Optional isolated 24Vdc @ 120mA output
 
Input/Output
Digital I/O: 6 or 12 relay outputs, Form A (normally open SPST contacts) rated at 200 Vdc @ 0.5A Max, 2 digital control inputs +5 to +12Vdc @ 20mA (optional), Control inputs may be used for record start/stop, alarm acknowledge and channel reset functions
 
Safety and Environmental
Operating Range: 0 °C to 50 °C, 10% to 80% RH non-condensing
Protection: IP65 when mounted in panel
Safety: Meets the requirements of EN61010-1 when installed in accourdance with the intructions in the manual.
UL and cUL: Pending

EMC: Meets the requirements of EN61326:2003 and CE directive 89/336/EEC.
Weight: Approximately 7 lbs. (3.17 kg) - weight will vary depending on options installed


Dimensional Drawings (click to enlarge)